Questions? +32 460 226 991 (English). EU-wide delivery!

Gas Detector for Refrigerants EVIKON E2630-HFC-230 R-410a

  • 5 variant

Gas Detector for Refrigerants EVIKON E2630-HFC-230 R-410a

EVIKON
  • Product code: E2630-HFC-230
  • Manufacturer code: E2630-HFC-230
  • Warranty: 2 years

Gas detector for R410a refrigerant EVIKON E2630-HFC, alarm signal: visual + audible, power supply: 90...265 VAC, measuring range: 0...500 ppm. More information

  • Availability

    Out of stock - can be requested

  • Out of stock  
Price excluding VAT € 201,00
Price with VAT 21 % € 243,21

Parameters and catalogs

Detector type Stationary
Equipment Siren 85dB, LED
IP Rating IP 65
Output signal 2x SPDT relay
Power Supply 230 VAC
Type of gas Refrigerant - R410a
Weight 0.39 kg

Declaration of conformity download

Download PDF

Catalog sheet download

Download PDF

Data sheet download

Download PDF

Installation instructions download

Download PDF

Catalog sheet eng download

Download PDF

The Evikon E2630-HFC is a detector designed to monitor the concentration of R410a refrigerant in air. The device uses a semiconductor sensor that provides accurate real-time measurements. Thanks to its reliable performance and easy installation, it is an ideal choice for commercial and industrial applications where refrigerant leak monitoring is essential.

The device is housed in a wall-mounted enclosure with IP65 protection, ensuring resistance to dust and water. The E2630-HFC detector is equipped with a visual and audible alarm that provides timely warning when the concentration of R410a rises. In addition, it offers two SPDT relay outputs for easy integration with ventilation control systems or alarms, ensuring a comprehensive safety solution.

Typical applications:

The E2630-HFC detector is intended for monitoring R410a refrigerant leaks in environments such as plant rooms, air conditioning and refrigeration equipment, refrigerant stores, or other installations where refrigerant concentrations need to be monitored. The device provides early leak detection, helping to minimise the risks associated with refrigerant leakage and protect both the environment and equipment.

R410a:

R410a is a synthetic refrigerant widely used in air conditioning and refrigeration systems. It is a colourless blend of two hydrofluorocarbon gases that ensures efficient heat transfer. Although it is non-flammable, leakage of R410a can contribute to greenhouse gas emissions and affect air quality. Monitoring R410a leaks is crucial for ensuring equipment safety and compliance with environmental standards.


Calibration certificate

Every Evikon gas detector package includes a factory calibration certificate. The calibration certificate is in English. Gas concentration measurement is affected by operating temperature and humidity. Therefore, calibration during detector production is carried out with enhanced safety in mind. 


Detector

Key features:

  • Detected substances: R410a refrigerant
  • May respond to: Other refrigerants R404a, R134a, R32, R-1234yf, R-1234ze, flammable gases, toluene, xylenes, vapours from cleaning agents, etc.
  • Sensor type: Metal oxide semiconductor sensor with diffusion sampling.
  • Measuring range: 0...500 ppm with 1 ppm resolution.
  • Two-stage detection
  • Response time: Less than 120 seconds with signal update every second.
  • Alarms: Flashing LED and siren (85 dB).
  • Recommended recalibration interval: 6–12 months
  • Expected sensor lifetime: > 5 years
  • Avoid strong mechanical shocks, vibration, or electromagnetic interference (EMI). Also avoid exposure to corrosive gases or products containing silicone.

Factory settings:

  • Calibration gas: R410a refrigerant
  • Measuring range: 0...500 ppm
  • Alarm settings:
    • ALARM1 = 200 ppm
    • ALARM2 = 500 ppm
  • Automatic alarm reset after ventilation (can be switched to manual mode using a magnet)

Specifications:

  • Power supply: 230 VAC
  • Power consumption: < 2 VA
  • Max. load of SPDT relay outputs: 250 VAC / 30 VDC, 5 A max
  • Operating conditions: -40...+70 °C, <95% RH, pressure 0.9...1.1 atm
  • Enclosure dimensions: H140 × W145 × D55 mm
  • Protection class: IP65

This detector is suitable for monitoring R410a refrigerant in a variety of environments, including industrial facilities and storage areas. It offers stable performance with low maintenance requirements, making it a practical solution for long-term use across a wide range of applications.


Register your detector

We will notify you of the statutory calibration due date.

  • Notification 1 month before calibration expiry
  • By email
  • Free, hassle-free service

Order recalibration

Recalibration is legally required at least once a year.

  • Mail-in calibration within 72 hours
  • On-site calibration
  • Calibration certificate

We respond to your product queries as quickly as possible, usually within 24 hours.

Nenašli jste to, co jste chtěli vědět?